[1]吴曦,刘建民,黄清海,等.第一代Matrix弹簧圈治疗颅内动脉瘤的安全性及中长期随访结果分析[J].介入放射学杂志,2008,(06):384.
 WU Xi,LIU Jianmin,HUANG Qinghai,et al.Middle and long-term follow-up of intracranial aneurysms treated with Matrix detachable coils[J].J Intervent Med,2008,(06):384.
点击复制

第一代Matrix弹簧圈治疗颅内动脉瘤的安全性及中长期随访结果分析()

PDF下载中关闭

《介入放射学杂志》[ISSN:1008-794X/CN:31-1796/R]

卷:
期数:
2008年06期
页码:
384
栏目:
神经介入
出版日期:
2008-06-15

文章信息/Info

Title:
Middle and long-term follow-up of intracranial aneurysms treated with Matrix detachable coils
作者:
吴曦刘建民黄清海
200433上海第二军医大学长海医院神经科中心
Author(s):
WU Xi LIU Jian-minHUANG Qing-haiXU Yi
Departm ent of Neurosurgery,Changhai Hospital of the Second Military Medical University,Shanghai200433,China
关键词:
Matrix弹簧圈颅内动脉瘤颅内支架
分类号:
R743.4
文献标志码:
A
摘要:
目的分析第一代Matrix可吸收聚合物涂层弹簧圈(简称Matrix)治疗颅内动脉瘤的安全性,及中、长期随访结果的影响因素。方法回顾性研究,本中心2003年5月至2005年7月75例使用Matrix栓塞颅内动脉瘤患者(79枚动脉瘤)。其中造影复查患者34例(35枚动脉瘤),平均随访时间12.5个月。根据复查结果对年龄、性别、动脉瘤最大径、瘤颈宽度、使用支架辅助、Matrix长度比例进行COX模型分析。结果Matrix相关的手术并发症率为13.3%,使用Matrix术后动脉瘤总复发率为11.4%,辅助支架治疗的大型动脉瘤的复发率为40%,未致密栓塞的动脉瘤中61.1%发生进一步栓塞。无统计学结果可证明年龄、性别、动脉瘤最大径、瘤颈宽度、支架辅助、Matrix占栓塞弹簧圈长度比例等因素影响Matrix栓塞动脉瘤的中长期预后。结论Matrix治疗颅内动脉瘤安全、有效,结合支架辅助技术可能会降低大型动脉瘤的复发率。Matrix弹簧圈有促进动脉瘤愈合的倾向,建议尽可能改善动脉瘤的血流动力学状况,此举将有利于动脉瘤长期愈后。Matrix弹簧圈的长期确切效果有待进一步随访观察及进行大样本的随机对照研究。

参考文献/References:

[1]Molyneux AJ ,Kerr RS,Yu LM,et al.International Subarachnoid Aneurysm Trial(ISAT)Collaborative Group.International subarachnoid aneurysm trial(ISAT)of neurosurgical clipping versus endovascular coiling in2143patients with ruptured intracranial aneurysms :a randomised comparison of effects on survival,dependency,seizures,rebleeding,subgroups,and aneurysm occlusion[J].Lancet,2005,366:809-817.
[2]Raymond J,Guilbert F,Weill A,et al.Long-term angiographic recurrences after selective endovascular treatment of aneurysms with detachable coils[J].Stroke,2003,34:1398-1403.
[3]Murayama Y,Nien YL,Duckwiler G,et al.Guglielmi detachable coil embolization of cerebral aneurysms :11years’experience[J].J Neurosurg,2003,98:959-966.
[4]Mitra D,Herwadkar A,Soh C,et al.Follow-Up of intracranial aneurysms treated with Matrix detachable coils:A single-center experience.AJ NR,2007,28:362-367.
[5]Taschner CA,Leclerc X,Rachdi H,et al.Matrix detachable coils for the endovascular treatment of intracranial aneurysms analysis of early angiographic and clinical outcomes.Stroke,2005,36:2176-2180.
[6]Szikora I,Seifert P,Hanzely Z,et al.Histopathologic evaluation of aneurysms treated with guglielmi detachable coils or matrix detachable microcoils.AJNR,2006,27:283-288.
[7]Meyers PM,Lavine SD,Fitzsimmons BF,et al.Chemical meningitis after cerebral aneurysm treatment using two second-generation aneurysm coils:report of two cases[J].Neurosurgery,2004,55:1222.
[8]赵振伟,邓剑平,高国栋.Matrix栓塞颅内动脉瘤的影像学随访[J].中华医学杂志,2007,87:37-40.
[9]Rivet DJ ,Moran CJ ,Mazumdar A,et al.Single-institution experience with matrix coils in the treatment of intracranial aneurysms :comparison with same-center outcomes with the use of platinum coils[J].AJNR,2007,28:1736-1742.
[10]Taschner CA,Leclerc X,Rachdi H.Matrix detachable coils for the endovascular treatment of intracranial aneurysms analysis of eaarly angiographic and clinical outcomes[J].Stroke,2005,36:2176-2180.
[11]Linfante I,Akkawi NM,Perlow A,et al.Polyglycolide /polylactide-coated platinum coils for patients with ruptured and unruptured cerebral aneurysms.A single-center experience[J].Stroke,2005,36:1948-1953.
[12]Fiorella D,Albuquerque FC,McDougall CG.Durability ofaneurysm embolization with Matrix detachable coils[J].Neurosurgery,2006,58:51-59.
[13]Niimi Y,Song J,Madrid M,et al.Endosaccular treatment of intracranial aneurysms using matrix coils early experience and midterm follow-up[J].Stroke2006,37:1028-1032.
[14]Murayama Y,Vinuela F,ISHII A,et al.Initial clinical experience with Matrix detachable coils for the treatment of intracranial aneurysms [J].J Neurosurg,2006,105:192-199.
[15]Wong GKC,FRCS(SN),Yu SCH,et al.Clinical and angiographic outcome of intracranial aneurysms treated with Matrix detachable coils in Chinese patients[J].Surg Neurol,2007,67:122-126.
[16]Jonas KT,Willinsky RA,Spears J,et al.Endovascular treatment of intracranial aneurysms with matrix coils:immediate post-treatment results,clinical outcome and follow-up[J].Neuroradiology,2007,49:223-229.
[17]Ding YH,Dai DY,Lewis DA,et al.Angiographic and histologic analysis of experimental aneurysms embolized with platinum coils,matrix,and hydrocoil[J].AJ NR,2005,26:1757-1763.
[18]Murayama Y,Vinuela F,Tateshima S,et al.Bioabsorbable polymeric material coils for embolization of intracranial aneurysms a preliminary experimental study[J].J Neurosurg,2001,94:454-463.
[19]Pierot L,Leclerc X,Bonafe A,et al.Endovascular treatment of intracranial aneurysms with matrix detachable coils:midterm anatomic follow-up from a prospective multicenter registry[J].AJ NR,2008,29:57-61.

相似文献/References:

[1]黄正松,戴钦舜.血管内治疗颅内动脉瘤[J].介入放射学杂志,1994,(02):104.
[2]华续明 万亮 江峰 刘鸣 马杰.TRUFILL DCS Orbit 水解脱弹簧圈栓塞治疗颅内小动脉瘤及宽颈动脉瘤[J].介入放射学杂志,2005,(05):532.
 HUA Xu-ming WAN Liang JIANG Feng LIU Ming MA Jie.Endovascular treatment of small and wide-nick intracranial aneurysms with TRUFILL DCS Orbit coil[J].J Intervent Med,2005,(06):532.
[3]陈左权,顾斌贤,张桂运,等.双C臂三维DSA在颅内动脉瘤栓塞治疗中的应用[J].介入放射学杂志,2006,(01):2.
 CHEN Zuo-quan,GU Bin-xian,ZHANG Gui-yun,et al.The utility of three dimentional DSA with bi-C-typed arms in the embolization of intracranial aneurysms[J].J Intervent Med,2006,(06):2.
[4]高不郎,李明华.使用液体栓塞剂Onyx治疗颅内动脉瘤[J].介入放射学杂志,2006,(02):118.
 GAO Bu-lang,LI Ming-hua.The embolization of intracranial aneurysms using the liquid embolic material of Onyx[J].J Intervent Med,2006,(06):118.
[5]宣井岗,彭亚,杨伊林,等.水解可脱弹簧圈栓塞治疗破裂急性期颅内动脉瘤的体会[J].介入放射学杂志,2006,(03):131.
 XUAN Jing-gang,PENG Ya,YANG Yi-lin,et al.Application of theTrufill detachable coil embolization in treatment of acute ruptured intracranial aneurysms[J].J Intervent Med,2006,(06):131.
[6]印建国,郭元星,郑曙光,等.颅内巨大动脉瘤栓塞一例[J].介入放射学杂志,2006,(04):256.
 YIN Jian-guo,GUO Yuan-xing,ZHENG Shu-guang.Embolization of giant intracranial aneurysm (one case report)[J].J Intervent Med,2006,(06):256.
[7]虞希祥,张顺开,司同国,等.颅内动脉瘤血管内栓塞治疗及并发症分析[J].介入放射学杂志,2007,(04):219.
 YU Xi-xiang,ZHANG Shun-kai,SI Tong-guo,et al.Intravascular embotherapy of intracranial aneurysm and analysis of complication[J].J Intervent Med,2007,(06):219.
[8]张纪庆,王志刚.水凝胶水解式微弹簧圈在颅内动脉瘤治疗中的应用[J].介入放射学杂志,2007,(10):710.
 ZHANG Ji-qing,WANG Zhi-gang.Hydrocoil for the treatment of intracranial aneurysms[J].J Intervent Med,2007,(06):710.
[9]杨志刚,刘建民,许奕,等.新型可回撤自膨胀支架在颅内动脉瘤治疗中的应用[J].介入放射学杂志,2008,(07):459.
 YANG Zhigang,LIU Jianmin,XU Yi,et al.Application of a novel retrievable self-expanding stent in intracranial aneurysm treatment[J].J Intervent Med,2008,(06):459.
[10]潘奇,刘建民,许奕,等.颈内动脉前壁动脉瘤支架结合弹簧圈栓塞术后再出血:病例报道[J].介入放射学杂志,2008,(08):560.
 PAN Qi,LIU Jianmin,XU Yi,et al.Rebleeding after stenting and coil embolization of an aneurysm in anterior wall of the internal carotid artery[J].J Intervent Med,2008,(06):560.
[11]吴曦,刘建民.第一代Martix可吸收聚合物弹簧圈治疗颅内动脉瘤的研制及应用现状[J].介入放射学杂志,2008,(11):828.
 WU Xi,LIU Jianmin.The present development and status of the first generation of Matrix bioabsorbable poly-meric material coil in the treatment of intracranial aneurysms[J].J Intervent Med,2008,(06):828.

备注/Memo

备注/Memo:
收稿日期:2008-02-04
更新日期/Last Update: 2008-06-15